Rotary evaporator motor anomaly solution

During the use of the rotary evaporator, sometimes the motor always has a high speed and there is no way to adjust the speed. The abnormality of this motor generally has the following reasons:

50 L rotary vacuum flash evaporator

1.The speed measuring generator coil is broken, and it can be tested by universal meter. The normal resistance is about 400Ω.
2.The generator coil is shorted to the housing.
3.The generator magnetic ring does not turn.
4. The generator magnetic ring is broken.

re-201d rotary vacuum flash evaporator price

After finding the cause, the solution has the following types:
1.If the generator is disconnected, replace the coil or switch on the trip point.
2.If the generator coil is short-circuited to the outer casing, open the generator rear cover of the motor to remove the open circuit.
3.Tighten the M5 nut.
4.Replace the generator magnetic ring.

10L rotary evaporator flask cost

The rotary evaporators produced and supplied by zzkd are divided into two types according to the capacity of the rotating bottle. They can be divided into two types: mini rotary evaporator and large scale rotary evaporator. Mini type has 2l rotary evaporator, 5l rotary evaporator, etc., and large ones have 10l rotary evaporator, 50l rotary evaporator and so on.